Korean Journal of Pediatrics ; : 463-469, 2011.
Article in English | WPRIM | ID: wpr-139022

ABSTRACT

PURPOSE: To validate the Dinamap ProCare 200 blood pressure (BP) monitor against a mercury sphygmomanometer in children 7 to 18 years old in accordance with the 2010 International Protocol of European Society of Hypertension (ESH-IP2) and the British Hypertension Society (BHS) protocol. METHODS: Forty-five children were recruited for the study. A validation procedure was performed following the protocol based on the ESH-IP2 and BHS protocols for children and adolescents. Each subject underwent 7 sequential BP measurements alternatively with a mercury sphygmomanometer and the test device by trained nurses. The results were analyzed according to the validation criteria of ESH-IP2. RESULTS: The mean (+/-SD) difference in the absolute BP values between test device and mercury sphygmomanometer readings was 1.85+/-1.65 mmHg for systolic BP (SBP) and 4.41+/-3.53 mmHg for diastolic BP (DBP). These results fulfilled the Association for the Advancement of Medical Instrumentation criterion of a mean+/-SD below 5+/-8 mmHg for both SBP and DBP. The percentages of test device-observer mercury sphygmomanometer BP differences within 5, 10, and 15 mmHg were 96%, 100%, and 100% for SBP, and 69%, 92%, and 100% for DBP, respectively, in the part 1 analysis; both SBP and DBP passed the part 1 criteria. In the part 2 analysis, SBP passed the criteria but DBP failed. CONCLUSION: Although the Dinamap ProCare 200 BP monitor failed an adapted ESH-IP2, SBP passed. When comparing BP readings measured by oscillometers and mercury sphygmomanometers, one has to consider the differences between them, particularly in DBP, because DBP can be underestimated.


Subject(s)
Adolescent , Child , Humans , Arm , Blood Pressure , Blood Pressure Monitors , Hypertension , Organothiophosphorus Compounds , Reading , Sphygmomanometers
2.
Korean Journal of Pediatrics ; : 463-469, 2011.
Article in English | WPRIM | ID: wpr-139019

ABSTRACT

PURPOSE: To validate the Dinamap ProCare 200 blood pressure (BP) monitor against a mercury sphygmomanometer in children 7 to 18 years old in accordance with the 2010 International Protocol of European Society of Hypertension (ESH-IP2) and the British Hypertension Society (BHS) protocol. METHODS: Forty-five children were recruited for the study. A validation procedure was performed following the protocol based on the ESH-IP2 and BHS protocols for children and adolescents. Each subject underwent 7 sequential BP measurements alternatively with a mercury sphygmomanometer and the test device by trained nurses. The results were analyzed according to the validation criteria of ESH-IP2. RESULTS: The mean (+/-SD) difference in the absolute BP values between test device and mercury sphygmomanometer readings was 1.85+/-1.65 mmHg for systolic BP (SBP) and 4.41+/-3.53 mmHg for diastolic BP (DBP). These results fulfilled the Association for the Advancement of Medical Instrumentation criterion of a mean+/-SD below 5+/-8 mmHg for both SBP and DBP. The percentages of test device-observer mercury sphygmomanometer BP differences within 5, 10, and 15 mmHg were 96%, 100%, and 100% for SBP, and 69%, 92%, and 100% for DBP, respectively, in the part 1 analysis; both SBP and DBP passed the part 1 criteria. In the part 2 analysis, SBP passed the criteria but DBP failed. CONCLUSION: Although the Dinamap ProCare 200 BP monitor failed an adapted ESH-IP2, SBP passed. When comparing BP readings measured by oscillometers and mercury sphygmomanometers, one has to consider the differences between them, particularly in DBP, because DBP can be underestimated.


Subject(s)
Adolescent , Child , Humans , Arm , Blood Pressure , Blood Pressure Monitors , Hypertension , Organothiophosphorus Compounds , Reading , Sphygmomanometers